Fat Suppressive and Susceptibility Tolerant PINS Pulses for Multi-Band Diffusion Weighted EPI

Eric Aboussouan1, Rafael OHalloran1, Anh Tu Van1, Samantha J. Holdsworth1, William A. Grissom2, Roland Bammer1

1Radiology, Stanford University, Stanford, CA, United States; 2Biomedical Engineering, Vanderbilt University, Nashville, Tennesse, United States

Multi-band acquisitions make diffusion imaging with full brain coverage faster and more SNR-efficient by simultaneously acquiring multiple slices with limited g-factor penalty. To do so, special RF pulses must be employed. PINS pulses excite multiple slices simultaneously with limited power deposition and can additionally be designed to be inherently fat suppressive. In this work, we present and compare two multiband excitation and refocusing schemes for Diffusion Weighted EPI (DW-EPI). The first method performs lipid suppression by shifting the fat slice in opposite directions during the excitation and refocusing steps. The second method reduces susceptibility artifacts by matching the time-averages of the slice selection gradients of the excitation and refocusing pulses.

Keywords

acquisition acquisitions adoption albeit amplitude amplitudes artifacts band bands behave behavior biomedical blipped blips blue bottom bound brain butts clearly clinical column completely complex conjunction conspicuous contrast cost coverage dark dashed deformation deformed designed dictate diffusion distortions downside eliminates employed exact excitation excite excited facto fashion field fields foundation frequency full fully gaps gradient gradients gray halving hard healthy homogeneous identical illustrate imaginary improper in vivo incomplete inhomogeneities inhomogeneity inhomogeneous inversion inverted issue lead legend light limited limits lipid long loss magnetization match matching matrix middle must need obviates occurs opposite outside overall pins positive possibly power precede preparation presence profile profiles protons pulse pulses radiology readout real recovery reduces refocus refocused refocusing relatively replicas requires ring saturation schemes selecting selection selective sensitivity separated separation series simultaneously since sinuses slice slices space spacing special steps suffers suggested suppression suppressive susceptibility taken technology temporal tilting tolerant traversed triangular turn utilize view visible water whole