Abstract #1847

A Comparison of MR-Tracking Methods in the Presence of Severe Metallic Artifacts and Physiological Motion

Wei Wang1, 2, Zion Tse3, Ravi T. Seethamraju4, Charles L. Dumoulin5, Tina Kapur1, Akila N. Viswanathan2, Robert A. Cormack2, Ehud J. Schmidt1

1Radiology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States; 2Radiation Oncology, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, United States; 3Engineering, University of Georgia, Athens, GA, United States; 4MR R&D, Siemens Healthcare, Boston, MA, United States; 5Radiology, Cincinnati Children's Hospital, Cincinnati, OH, United States

The presence of metal in MR-guided interventional procedures affects MR-tracking quality by causing static magnetic field and radiofrequency field inhomogeneities. A custom-built carbon fiber catheter for radiation bracytherapy applications was tested using different MR-tracking methods (zero-phase reference and Hadamard multiplexing schemes without or with phase-filed dithering (PFD) strategy). The test environment included surrounding metallic needles and simulated physiological motions from a motional platform. Results showed an MR-tracking capability at 40 fps and 0.6 &[mult] 0.6 &[mult] 0.6 mm3 and demonstrated that the Hadamard scheme produced a more accurate and robust result than zero-phase-reference, and the use of PFD dramatically enhanced both tracking schemes.

Keywords

absence accuracy accurate adding addition affects allowing amplitude antennas applies around artifacts assembly assigned attached away black built capability carbon catheter causing cervical channel children coil coils conditions conductive conductivity consisted control curve custom cycle cycles described detailed device diamagnetic displacement distorting dithering efficient enabled entire environment erroneous error experimental fiber field five flash frequencies frequency gold gradients guided gynecologic highest hospital improved incorporated inhomogeneities inserted integrated interstitial length lengths limitation long makes measured medical metal metallic motion motional multiplexing needle needles none object orientation orthogonal oscillation oscillatory pattern peak phantom physiological platform position positional positioning positions practical precisely presence procedure produced quality radiation radiology rapid recording reduction reliability removed required resolution respiratory rest robust robustness routinely scheme schemes school select selects sets setup several severe sharpest simulating solenoid spatial speed static strong strongest sufficient suitable surrounding susceptibility system template therapeutic tissue tracked tracking translational treatment tumors variable vector women zero