Using a Simple Thermal Model to Correct Errors in PRFS MR Thermometry Due to Heat-Induced Tissue Susceptibility Changes

Paul Baron1, Roel Deckers1, Martijn de Greef1, Job G. Bouwman1, Chris J.G. Bakker1, Lambertus W. Bartels1

1University Medical Center Utrecht, Utrecht, Netherlands

During MR-guided high intensity focused ultrasound (MR-HIFU) ablation of fat-containing tissues like the human breast or liver, heat induced susceptibility changes may give rise to local field disturbances, leading to errors in PRFS thermometry of the water-based tissues. A correction method is presented that combines the initial uncorrected measured temperature change with a thermal model to estimate the field disturbance and thus correct for these errors. The feasibility is shown in a porcine adipose tissue and ethylene glycol gel phantom. The maximum temperature error decreased from 3C before correction to less than 1C after the correction.

Keywords

ablation absolute acoustic actual additionally adipose agar allowed although amplitude application applying baron beam breast cell clinical close coefficients coil combined complex consisted constant containing convert correct corrected correction cylinder defined degree dependent determine deuterium diagram diameter distributions disturbance disturbances domain dynamically echoes edge efficient electron error errors ethylene experiment experimental explained fast feasibility field fits fitting flow focus form frequency function give glycol gradient guided half heat heating human improved induced initial input integrated intensity interface intersected iteration iterations known leading least liver local located maps matrix measured medical memory minimized mixed model nine normalized nuclear overview peak peaks perpendicular phantom porcine position positioned post power principle procedure processing profiles real recalled receive reduced remains represented respectively screening separate separated separation setup silica simple since slice sons spacing spoiled square stabilize stable starting step steps sufficient susceptibility system temperature thermal thermometry tissue tissues types uncorrected variable variables versus volumetric water