Abstract #1614

On the Reliability of in Vitro Mapping of Ultrashort T2* of Porcine Intervertebral Discs: A Detailed Analysis of the Signal Decay

Lena Altherr1, Stefan Kirsch1, M. L. R. Schwarz2, Gregor Reisig2, Dieter Schleich2, Lothar R. Schad1

1Computer Assisted Clinical Medicine, Heidelberg University, Mannheim, Germany; 2Department for experimental orthopaedics and trauma surgery, OUZ Mannheim, University Medical Centre, Heidelberg University, Mannheim, Germany

Mapping of ultrashort T2* with UTE sequences provides high SNR/time and should be well suited for studying degeneration of musculoskeletal tissue in vivo. However, chemical shift and magnetic field inhomogeneities can strongly influence the shape of the T2* signal decay. The reliability of T2* mapping performed on isolated porcine intervertebral discs was investigated with a standard 2D-UTE sequence at 9.4 T. Results show that the signal decay is affected regionally by different physical effects. Interpretation of T2* maps should therefore be handled with great care.

Keywords

acquisition adjoined affected alternatively analytical animal applicable application assisted axial axis calculating care cartilage caudal caused chemical chemically choice chosen clinical comparable complexity component components considerations consisted core cranial curve decay decays degeneration delay detailed dieter differ directly discs domain excitation exemplary experimental exponential faster field fits fitted forthcoming frequency fresh functions furthermore generated global goodness great handled in vivo increments influence inhomogeneities inner interestingly interpretation investigate isolated kirsch longer lumbar macroscopic making mapping maps match materials matrix measured measuring medical medicine middle might mode model modeling modulation mostly motion motivated musculature nucleus often operating oriented outer paper parallel particular performance phenomena physical pixel porcine preparation prevent problem processes projections prone pulse radiology refocusing relaxing reliability removed resonator ring sample samples satisfactory scanner segments shape side slice slower soft spinal spine started still strong strongly studies suggests suited superposition surgery surrounding theoretical tissue transverse trimmed ultrashort variable vertebrae visualize vitro whole wise yield