Bo Xu1, 2,
Tian Liu2, Pascal Spincemaille2, Yi Wang1,
2
1Biomedical
Engineering, Cornell University, Ithaca, NY, United States; 2Weill
Cornell Medical College, New York, United States
Flow artifacts in quantitative susceptibility mapping (QSM) are corrected with a fully flow compensated multi-echo gradient echo sequence combined with a hybrid linear-quadratic fitting of the phase data. Phantom experiments and in vivo studies demonstrate improvements in the estimation of the field inhomogeneity and the susceptibility map. This enables more reliable venous oxygenation quantification in the brain with QSM.