A Lightweight, Portable MRI Brain Scanner Based on a Rotating Halbach Magnet

Clarissa Zimmerman Cooley1, 2, Jason P. Stockmann2, 3, Brandon Dean Armstrong2, 3, Matthew S. Rosen2, 3, Lawrence L. Wald2

1Electrical Engineering, Massachusetts Institute of Technology, Cambridge, MA, United States; 2Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States; 3Department of Physics, Harvard University, Cambridge, MA, United States

A lightweight brain scanner based on a rotating permanent-magnet Halbach array is constructed and tested in phantom imaging. As the magnet is rotated around the object, the multipolar static field inhomogeneity creates generalized projections. The image is reconstructed using an iterative algebraic reconstruction technique based on knowledge of the field map. We show 2D phantom images containing aliasing artifacts similar to radial PatLoc images, which share the multipolar encoding field. These artifacts are greatly reduced through parallel imaging methods, which further reduce streaking and other residual artifacts while improving resolution.

Keywords

accelerate achieved address adds another arena around array audience authors axis ball balls bandwidth basis bedside benefit biomedical blurring brain built calculating calibration capillary careful clearly coil coils complications concept consisted corrections cover creating cylinder dean decrease directly doped dots drift eighty electrical eliminates emergency encoding engineering engineers example excitation expected experiment experimental exposes external field fields fixed functions gradient ground half help hemorrhage hyperbolic implemented improved improvements includes inconveniences inhomogeneity inhomogeneous initial instr knowledge lightweight loop lowest magnet maps materials matrix measured monitoring necessary noise novel object obvious permanent phantom polynomial portability portable power prior probe probes projection projections proof pulse pulses radiology readout receive reconstructed reconstruction refocusing regulated requiring rotated rotating rotation rotations roughly scanner sensitivities sensitivity shape sharp shielding simulated slice solenoid space spatial spin spins stroke sufficient suitable sulfate supplies supported surface switched symmetry system systems target temperature throughout traumatic truth unique unlike variation version volume water