Analysis of Thermal Stability of Background Phase Errors in Phase-Contrast Flow Imaging

Julia Busch1, Signe Johanna Vannesjo1, Christoph Barmet1, 2, Klaas P. Pruessmann1, Sebastian Kozerke1, 3

1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland; 2Skope Magnetic Resonance Technologies, Zurich, Switzerland; 3Division of Imaging Sciences and Biomedical Engineering, Kings College London, London, United Kingdom

Background phase errors of various spatial and temporal orders limit the accuracy of phase-contrast magnetic resonance imaging. Current calibration techniques can compensate for them almost completely; however, they require the MR system to be temporally stable. In the present work we analyze the stability of background phase errors under thermal changes of the gradient system as it may occur for scans of long duration and high duty-cycle. Measurement of the gradient impulse response function shows a shift in frequency of the oscillatory field fluctuations with increasing temperature which results in a change of 0th and 1st order field offsets.

Keywords

accuracy addressed affected amplitude analytical analyze analyzed applied approaches arising aspects assuming axes axis background besides best biomedical blood calculation calibration camera cardiac chain choice coil coils college compromising concern concomitant conditions considering constant contrast corrected correction corrections cycle dependence depending determined development directional disabled division domain duration duty dynamic eddy effective emphasis encoded encoding engineering ensure epoxy error errors fiber field flow fluctuations frequencies frequency function functions gatehouse gradient heart impulse increasing institute king lead limited linear longer measured mechanical might minor model modify monitored monitoring mount nevertheless objective occur offsets onto optic oscillations oscillatory periods phantom phantoms prerequisite proposed recently recorded remain render reported reproduced reproducible residual resolution response retrospective room scanning sciences separately sessions setup short since slice software spatial stability stable static stationary steps still subtraction sufficient suggested support system technologies technology temperature temperatures temporal term thermal thermally thermo triggering unchanged upon upper validate various velocity view