Abstract #4162

Reduction of Eddy-Current-Induced Distortion in Diffusion-Weighted EPI Using Spin-Tagging

Kun Zhou1, Wei Liu1

1Siemens (Shenzhen) Magnetic Resonance Ltd., Shenzhen, Guangdong, China

A spin-tagging based method is proposed to correct eddy currents induced distortions in DW-EPI images. In this method strips placed by tagging pulses are used to characterize image distortion. Parameters (scaling and shifting factors) describing distortion are derived by comparing strip images of distorted image and reference image (image with b = 0). After that pixel shifting map is calculated using these parameters and used to perform distortion correction.

Keywords

abstract academic acquisition aligned appears applied artifacts bandwidth becomes blurring calculation certain characterize china column columns considered consists contain contrast correct corrected correction correlation depends described describing determine developed diffusion directly distorted distortion distortions domain eddy efficient encoding erroneous fits fitted fitting frequencies frequency function generated gradients handbook hence ideally implemented improve improving inconstant indicates individual induced introduced isotropic lead lies linear linearly located magnify maps matrix measure measured mode notation note obvious original orthogonal overcome peak people phantom pixel placed planar played position post precise preparation prepared press problem procedure processing produce proposed pulse pulses quality random rapid readout reduce reduced reduction remove representation republic require required respectively saving scaled scaling scanner scanners scheme selection slice space spin step steps strip stripes strips subtracting switching tagged tagging translate translating uncorrected validate variations