Abstract #4082

Effects of in vivo Flip Angle Variations on the Accuracy of DCE-MRI Perfusion Parameters at 1.5T and 3.0T

Jiangsheng Yu1, Xia Zhao1, Yiqun Xue1, Mark A. Rosen1, Christina S. Chu2, Hee Kwon Song1

1Departments of Radiology, University of Pennsylvania, Philadelphia, PA, United States; 2Obstetrics and Gynecology, University of Pennsylvania, Philadelphia, PA, United States

The goal of this work is to determine to what extent flip angle deviations affect DCE-MRI perfusion parameter accuracy on three different MRI systems, including whether higher fields and short bore systems can potentially exacerbate these measurement inaccuracies. The results show that over a large region of the body, the ranges of flip angle deviations are similar for the three systems. These variations can potentially mask true changes in tumor treatment response, and flip angle mapping may be a crucial in DCE-MRI protocols.

Keywords

accuracy accurately acknowledgments acquisition actual adjusted advantage affected agents amplifier anatomic angular applied arterial assume assumed axial bandwidth blood body bolus bore calibration cancer cause channels close combined complete compute computed condition contrast coronal crucial degrees determine determined deviation deviations differ dynamic effectiveness emerged enhanced error errors every exacerbated exacerbates expected extent fields finite fitted fitting form fractional function functional gain generate gradient gynecology hand hardware homogeneity identical in vivo inferior input intravenous intrinsic lateral lesions limitations located magnetizations maps mark mask middle model modified moment nearly next nominal noted obstetrics orientation orientations oversampling parker part particular perfusion peripherally pixel played population potential potentially prescribed previously prime procedure protocol proximity radial radiology random readout remain remained repeat respectively response scanner series short significantly simulation slice slices smallest sonata song stable studies subsequently substantial substantially suggests superior system systems table take throughout tissue trans treatment trio true tumor tumors utilized variations views wavelength whether