Phase-Derived Vascular Input Functions for 2D DCE-MRI of Cerebral Gliomas: Reproducibility and Diagnostic Value

Greg O. Cron1, 2, Thanh B. Nguyen1, 2, Rebecca E. Thornhill1, 2, Jean-Francois Mercier1, Claire Foottit1, Carlos H. Torres1, 2, Santanu Chakraborty<s

1The Ottawa Hospital, Ottawa, ON, Canada; 2The University of Ottawa, Ottawa, ON, Canada

For DCE-MRI of cerebral gliomas, temporal resolution can be increased using 2D sequences with limited coverage. However, the resultant saturation and inflow effects distort the measured vascular input function (VIF). A relatively new way to solve this problem is phase-derived VIFs (VIFφ). The purpose of this study was to compare the reproducibility and diagnostic value of tracer kinetic parameters (TKPs) calculated with individually-measured VIFφ and a published population-averaged VIF (VIFpop), in a group of 31 patients. VIFφ was superior to VIFpop for TKP reproducibility. VIFφ- and VIFpop- derived TKPs were equally good at distinguishing low- from high- grade tumors.

Keywords

accepted accounting adversely affect aggressiveness analyses assuming axial blood bookend bottom cancer cerebral coefficient cohort concentration concentrations contrast correlation coverage curve derived diagnostic diagnostically distortion distribution dose dynamic employing enhanced equal evaluated flash function functions grade grading hogan hospital hypothesized improves inaccuracies increasing inflow input intraclass introduce jean john kinetic least limited magnitude maps mark measured muscle must neck overall parker patient patients population possibility post predicting prediction problem published quantitative relatively reliability reproducibility reproducible resolution saturate saturation saved severe since sinus slice slices solve statistic strategy studied superior surgery table temporal tracer tumor underwent upper useful valuable variable variation vascular versus vessel volume vulnerable whether