Abstract #3489

A Setup for Continuous Arterial Spin Labeling with a 4 Channel Radiative Labeling Coil Allowing for High Duty Cycle Labeling at 7T

Wouter Koning1, Johanneke H. Bluemink1, Esben Petersen1, Alexander Raaijmakers1, Anke Henning2, Cornelis A.T. van den Berg1, Jaco Zwanenburg1, Peter Luijten1, Dennis W

1University Medical Center, Utrecht, Netherlands; 2ETH Zurich, Zurich, Switzerland

A setup is presented for the use of radiative antennas as external labeling coils for cASL at 7T. Four radiative antennas attached to a neck pillow filled with heavy water are used to create a B1 field in the neck. This neck pillow can be used in combination with a head coil for labeling during ASL experiments. SAR calculations show that using this setup labeling pulses for cASL can be used with a factor 5.5 longer duration than is allowed with the use of the head coil, while remaining within SAR safety guidelines.

Keywords

absorption achieved acknowledgments acquisitions acting aimed allow allowed allowing alone amplitude antenna applied array arterial assuming available avoids becomes benefit berg better birdcage blood body carotids causing cerebral channel coil coils combination combined comes compromises consisting continuous control corresponds coverage cycle depending dielectric driven duke duration durations duty effectiveness efficiency element entire especially eventually exceeding experiment experimental external facilitates family field fitting flow focusing foundation gain going gradient guidelines head heart heating height henning homogenizing human in vivo input inside jaws label labeled labeling larger limit limited limiting local long male maps materials measured medical medicine model modes molecular neck needed neglected normalized nova overview peak period peter pillow place potential power procedure pulse pulsed pulses radiative receive receiver reduces respectively restriction safety setup shim simulations since slice slices spin strengths strongly sufficient supported systems table takes tight translational transmission transmit transverse traveling twofold usually visualize volume volumes watt wave waveguide whereas