Abstract #3430

Combining Active Markers and Optical Tracking for Prospective Head Motion Correction

Murat Aksoy1, Melvyn Ooi1, Ronald D. Watkins, Daniel Kopeinigg, Christoph Forman2, Roland Bammer1

1Center for Quantitative Neuroimaging, Department of Radiology, Stanford University, Stanford, CA, United States; 2Computer Science, Friedrich-Alexander-University Erlangen-Nuremberg, Erlangen, Germany

Active markers and optical tracking systems (i.e. cameras) have been used for prospective correction of motion artifacts in head MRI. Optical tracking systems have the advantage that they require no additional MR data acquisition (i.e. navigators) to detect motion, whereas active markers can detect motion without the need to perform cross-calibration between the scanner and the tracking system. In this study, we combined active marker and optical tracking to benefit from the advantages of both systems. The active marker system was used to perform cross-calibration of the optical tracking system in a very short time with no discomfort to the patient so that the actual tracking can be done with the camera.

Keywords

accomplished accuracy acid active additional advantage advantages agar already anal artifacts assembly assessed battery beginning black calibration camera cameras checkerboard clinical coil coils combined combining computer consist constant corrected correcting correction cross decrease defining detect detection determine displaced done dotted drilled edges emanate enough error errors established every exam examination executed existing expected explained feasibility fiber field foundation frame function gave geometrical geometry glass gradient head holes important improve independent inhomogeneities initial inside interference intervention known landmark link long marker markers motion mounted navigator needs negative nonlinearities nonlinearity optic optical part patient pattern phantom placements position positions positive powered practice preliminary problem projection proposed prospective quantitative radiology reaching readout readouts reduce relation relationship remain remaining require respect rigidity rigidly rotations running scanner scanning setting setup short simulate since specifically stage stages step successfully supported system systems table tedious tend thereafter towards tracking translations transmission unlike unmounted user video view whether whole