Abstract #3426

Phase Correction for Interslice Discontinuities in Multislice EPI MR Elastography

Matthew C. Murphy1, John Huston, III1, Kevin J. Glaser1, Armando Manduca2, Joel P. Felmlee1, Richard L. Ehman1

1Department of Radiology, Mayo Clinic, Rochester, MN, United States; 2Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, United States

MR elastography (MRE) is an MR technique for noninvasively measuring tissue stiffness. Some applications, such as the brain, require a volumetric acquisition and 3D inversion because of through-plane wave propagation. An EPI pulse sequence is useful for such an application because it can acquire a volume of data with short acquisition times. However, slice-to-slice discontinuities in the phase of the EPI images can arise will bias the final inversion results. The purpose of this work is to present a filtering method that removes these slice-to-slice discontinuities, and as a result improves MRE repeatability.

Keywords

acceleration accurate acquisition amplified application applications arise artifacts assess asset avoid axial beginning bias biomedical bottom brain calculating chosen clinic coefficient collected combined complex component conjugate constant contain contrast coronal correction create curl days determined direct discontinuities displacement driver encoding engineering exam example exams excite external filter filtered filtering final free frequency function global gradient gradients head improved improves inspected introduced inversion inversions inverted isolating john leading like mathematically matrix mayo measuring middle modified motion must necessary negative offsets original pass period physiology pillow placed positive process processed produce produced product propagation proposed protocols pulse radiology reconstructed record rectangular reduced refocusing reformatted removes repeatability repeated represents require science shear short side significantly slice slices slowly soft source space step stiffness subject synchronized tissue true typically underestimate useful variation varying vibration volume volumetric wave waves window worst