Rapid and Low SAR B1-Mapping Using a BURST-Based Bloch-Siegert-Shift Sequence

Alexander Gotschy1, 2, Uvo C. Hlscher3, Thomas C. Basse-Lsebrink1, 4, Andr Fischer5, Morwan Choli3, Thomas Kampf1, Volker Sturm1, Daniel Neumann3

1Department of Experimental Physics 5, University of Wrzburg, Wrzburg, Germany; 2Department of Internal Medicine I, University of Wrzburg, Wrzburg, Germany; 3Research Center Magnetic Resonance Bavaria (MRB), Wrzburg, Germany; 4Department of Neurology, University of Wrzburg, Wrzburg, Germany; 5Institute of Radiology, University of Wrzburg, Wrzburg, Germany

An accurate knowledge of the B1+-field distribution becomes increasingly important for many MR applications like B1+ shimming, Spatially Selective Excitation algorithms or guaranteeing patient safety at ultrahigh fields. To meet the requirements of high field applications we used the lately introduced Bloch-Siegert-Shift based B1+-mapping method in a BURST sequence. The new sequence combines robustness against T2* effects with reduced SAR. In the presented study we found that a centric encoded BS-BURST sequence allows highly accelerated B1+-mapping with superior quality. The B1+ map of a whole head could be acquired in 36s which is adequate for applications in clinical routine.

Keywords

accelerated accordance achieves acquisition additional adjusted adjustments agreement allow allows already amplified applications applied artifacts artificially assessed becomes brain burst calculation centric clinical combine consent consisting constant contrast cycling deliver detection determination determined deviation diagram displayed displays distribution doped double duration enable encode encoded encoding essential evaluated excitation experiment experimental extended extinction field fields five fivefold frequency furthermore good gradient guaranteeing guidelines hardly head healthy highly human important in vivo increasing increasingly independent informed inhomogeneous institute institutional interleaved internal introduced knowledge like linear magma magnitude makes many mapping maps matrix meet neurology noisier optimize outperformed outperforms particularly patient peter phantom physics picture pineapple please project proposed pulse pulses quality radiology rapid reduced refer refocusing regarding regular requirements robustness routine safety saturated scanner schema selective sensitive shaped shimming significantly simplified slice slices slightly spatially spin stability stayed steps still strength suitable superior supported termed thereby turbo ultrafast ultrahigh valuable volunteer water whereas whole wise