Abstract #2860

Comparison of Simultaneous Multiband Whole Brain Imaging with Multiplexed-EPI

Liyong Chen1, 2, Audrey Chang1, 2, Junqian Xu3, Steen Moeller3, Edward J. Auerbach3, An Thanh Vu1, 2, Kamil Ugurbil3, Essa Yacoub3, David

1Helen Wills Neuroscience Institute, University of California, Berkeley, CA, United States; 2Advanced MRI Technologies, Sebastopol, CA, United States; 3Center for Magnetic Resonance Research (CMRR), Department of Radiology, University of Minnesota, Minneapolis, MN, United States

Multiplexed-EPI is performed using different number of images readout simultaneous in the EPI echo trains. Compared in whole brain imaging, slice accelerations (M) of 6 to 48 images per echo train were evaluated. Results showed observable trade-offs in image quality at highest accelerations using single echo trains to scan the whole brain and good quality images using two or three echo trains.

Keywords

acceleration accelerations achieve aliasing amplitude appeared array arrays axial bottom brain channel coil coils combination contribute controlled distortions encode equals excitation facto faster fewer fields funded gave general generally gradient grants greater halving head held hows human improve inserts institute issues larger length lengthened likely limit loss matrix maximize might minimized mitigated multiplex multiplexed needs oblique observable offs overage partial partially press process pulses quality radiology ramp ranging reached recently record reduce reduced reduction referred repeated required resurgence rows sacrifice sampling scanner separate shot slew slice slices spacing spacings speed subjects switching talk together trade train trains trio whole worsening