Abstract #2812

A 64-Channel Brain Array Coil for 3T Imaging

Boris Keil1, James N. Blau1, Philipp Hoecht2, Stephan Biber3, Michael Hamm3, Lawrence L. Wald1, 4

1A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, United States; 2Siemens, Medical Solutions USA, Charlestown, MA, United States; 3Siemens Healthcare, Erlangen, Germany; 4Harvard-MIT Division of Health Sciences and Technology, Cambridge, MA, United States

In brain imaging, parallel acquisition has impacted clinical applications to the point where nearly every brain examination is performed with an array comprising multiple smaller surface coil elements. In this study we design, construct, and evaluate a 64-channel brain array coil and compare it to a 32 channel coil, constructed with the same coil former geometry, in order to precisely isolate the benefit of the two-fold increase in array coil elements We validate the arrays with SNR and g-factor maps and accelerated in vivo imaging.

Keywords

ability accelerated accommodate acquisition acquisitions active addressed addresses allow anatomically anterior approximately arrangement array biomedical brain capacitor capacitors channel channels circles circuit circuitry circular clinical closes coax coil coils comfort components comprising conductor consisting construct constructed cortex critical daily decoupled decoupling design detected detuning disseminated division domain element elements every examination except facilitate finally fold former frequency gains general geometrically geometry halves head health helmet highly impacted impedance improved improvements in vivo inductance intermediate isolate isotropic joining larger location long loop loops maps matched matching material matrix measured medical multiplex must nearest nearly neck neighbor neighboring neighbors next noise onto open output outs overlap overlapping overlaps pair pairs parallel part patient patterns pentagonal performance peripheral portion posterior preamplifier pulled radiology rather receive respectively robust rounded scanner school section segment semi sensitivity series served shaped sharing slice slightly snap spin studies subject suited surface symmetrically system tapers tuning typical unit utilizes validated variable wide