A Volume Coil and a Multichannel Receive Array for High Resolution Hand/wrist Imaging at 7 T

Sairamesh Raghuraman1, Frank Resmer2, Matthias F. Mueller2, Michael Manglberger2, Stefan Zbyn3, Claudia Kronnerwetter3, Peter Baer4, Siegfried Trattnig3, Titus Lanz<su

1MRB Research Centre, Wuerzburg, Bayern, Germany; 2RAPID Biomedical GmbH, Rimpar, Germany; 3Dept of Radiology, MR Centre - High Field MR, Medical University of Vienna, Austria; 4Siemens Healthcare, Munich, Germany; 5Dept of Experimental Physics - V, University of Wuerzburg, Germany

High resolution MR imaging of hand is mandatory both to improve early diagnosis and for our understanding of diseases like Rheumatoid Arthritis and Osteoarthritis of wrist and finger joints. While 7 T promises excellent SNR for high resolution imaging of the hand, the usual approach of transmit or transceive arrays may not be the best solutions here, since arrays with a very high density and filling factor are required. This is due to the fact that the hand is a relatively small object and sample noise domination is not necessarily achieved. We present a combination of a volume transmit coil and a 12 channel high density receive (Rx) array for wrist imaging at 7 T.

Keywords

abdomen acceleration achieved acquisition actively allows anatomical around array arrays arthritis band best biomedical birdcage calibration capacitance caused channel coil coils combination configuration constructed constructing correlation curved decouple decoupled decoupling density dept details diagnosis diameter dielectric diodes diseases distance distort distributed domination drop electrical electrically electronics elements enable ensures excellent excitation experimental fact feet field fine finger finished frank frequency furthermore fuses good hand head healthy hence hold holder horizontally housing human hybrid improve incorporated individual infeasible initial inside instead joints legs length like load loaded located manageable mandatory manual medical metacarpal minutes nearly necessarily noise object operating optional osteoarthritis pass passive patient peter physics placed plate ports position positioned possibility progress properly quadrature radiation radiology rapid receive reception reduce reduces reduction reflection relatively resolution reveal rheumatoid sample scanner schematic shield shielded since spatial studies superman switch tolerated transmit traps tune ultra unlike usual vertically voltages volume volunteer water wrap wrist