An 8-Channel Dual-Tuned 1H/19F Flexible 7 Tesla Body Coil with Meanders

Stephan Orzada1, 2, Andreas K. Bitz1, 2, Mark E. Ladd1, 2, Kai Nassenstein2, Stefan Maderwald1

1Erwin L. Hahn Institute for MRI, Essen, NRW, Germany; 2Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, NRW, Germany

Ultra-high field strengths of 7 T and above promise a high SNR which is useful especially when trying to image nuclei with only a small incidence such as 19F, but inhomogeneities due to the short wavelength cause severe problems. In this work we present a flexible 8ch Tx/Rx coil for human abdominal imaging of 1H and 19F at 7 Tesla allowing the use of the same B1 shim for 1H and 19F. The array and the 8-ch transmit system were successfully tested ex vivo for both 1H and 19F.

Keywords

acquisition adding adjust ages allow already always amplifiers anders another applied around array arrays axial body built capacitance capacitor causes channel channels choosing chosen close coefficient coil comes contained contrast cope coronal correctly coupling dedicate detected detection diagnostic discuss double dropouts dual effective eight elem element elements elliptical emulsion enable ency especially even field flexible frequency future grad head heir horse hospital human inductance inside introduce introduced leads least length liver make maps match matching meanders metal micro need network notable nuclei optimized overlaid parallel phantom phasing plastic problem proposed proximity quired receive resolution respectively scale schematic sector series shim shimming short shorted since smaller solution still strengths strip success suitability switches system table though tolerances trying tuned ultra variant vector vial vibe view wavelength whole