RF Safety of Aneurysm Clips at 7 Tesla: Effect of Field Polarization

Yacine Noureddine1, 2, Oliver Kraff2, 3, Mark E. Ladd2, 3, Gregor Schaefers4, Karsten H. Wrede5, Andreas K. Bitz2, 3

1MR:comp GmbH, MR Safety Testing Laboratory , Gelsenkirchen, NRW, Germany; 2Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, NRW, Germany; 3Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, NRW, Germany; 4MR:comp GmbH, MR Safety Testing Laboratory, Gelsenkirchen, NRW, Germany; 5Clinic for Neurosurgery, University Hospital Essen, Essen, NRW, Germany

RF-induced heating as a result of electric field elevations in tissue close to metallic aneurysm clips is a major concern with respect to the safety of patients with these implants. In this study, a detailed investigation of the orientation of the elongated implant with respect to the polarization of the electric field was performed. Polarization and incident angle of a uniform plane wave were varied in steps of 90 degrees utilizing numerical simulations. Additionally, field measurements were performed and compared to the simulations.

Keywords

afterwards aligned allow alloy aneurysm angiography assessment avoided axis blood classified clinic clip clips close coil coiling coils comp comparable compliance computation considered constant contact demo depend dependency depiction design detailed determine difficult dimension dimensions direct distance distribution electric elev elevation elevations elongated ends establish examination excluded expected exposure field forces generated generation head heat hence highest homogeneously implant implants induced influence inner inside institute intervention invest kept laboratory liquid local located location longitudinal made magnitude mandatory mark materials meander mechanisms mesh million minutes model monitoring moreover must near need neurosurgery next noninvasive normalized numerical oliver orientation orthogonal overall parallel patients phantom placed polarization position post produces propagation quality radiation radiology random rather realistic reason reduce reduced reduces regarding resolution respect safe safety scenarios separately simulated simulating simulation simulations since space static steeps temp therapy though tissue torque transmit treated turned typical uniform unsafe utilized valuated varied vary wave whereas wise