Heating Near Deep Brain Stimulation (DBS) Lead Electrodes During Imaging with a 3T Transceive Head Coil in Cadaveric Porcine Heads

Devashish Shrivastava1, Timothy Hanson2, Ute Goerke1, Lance DelaBarre1, Rachana Visaria3, Paul Iaizzo1, Aviva Abosch1, JT Vaughan1

1University of Minnesota, Minneapolis, MN, United States; 2University of South Carolina; 3MR Safe Devices LLC, Minneapolis, MN, United States

Heating near DBS lead electrodes was measured using fluoroptic temperature probes and simulated using a combination of a proton resonance frequency (PRF) shift based MR thermometry method and bioheat thermal modeling in cadaveric porcine heads during imaging with a 3T transceive head coil. The effect on the heating of the placement of the extra-cranial DBS lead with respect to the coil was investigated to reduce the heating. The feasibility of using the MR thermometry and thermal modeling together as a hybrid approach on predicting maximum heating near the electrodes is demonstrated as an online tool to verify patient safety.

Keywords

ability absolute accuracy accurate animal around artifact artifacts avoid away axial back better blood brain clinical clinically closeness coil comparable concern conductive contacts contrast cranial damaging deep density design desired determine develop deviation device devices distal electro electrode electrodes engineering evaluate expected extra fact feasible field fields flux gradients harmful head heads health heating herein hybrid implantation in vivo influence inputs inside lance lead long loop maximize maximized measured mine minimize minimized model models near needed next odes oped organization outside parallel patient peak perfusion physiologic pieces place placed placement placements plausible porcine preliminary probe probes produce promise properties protocol protocols proton recalled recently reduced regularly relatively requires respect safe safety scalp serves significantly simulated simulating since south spatial spin stimulation studies susceptibility swine taped tech temperature temperatures temporal theoretical therm thermo thermometry tissue together tool turbo typical ultra validate validated validating verify viable waits whole world