Meeting Banner
Abstract #2702

MRI Enhancement Via High Dielectric Constant (εr = 510) Pad at 3T: Brain Imaging

Qing Yang1, 2, Chris Collins, Michael Lanagan3, Zachary Herse, Sebastian Rupprecht, Wei Luo, Yeun Chul Ryu, Megha Patel, Jeff Vesek

1The Penn State College of Medicine, Hershey, PA, United States; 2Neurosurgery, The Penn State College of Medicine, Hershey, PA, United States; 3The Penn State University

High dielectric constant (HDC) pads provide a simple, effective, low-cost method for improving the quality and safety of MRI in a number of applications at 3T. Our results indicate that when HDC-pads (filled with barium titanate (BaTiO3)and deuterium oxide (D2O) are placed bilaterally along the sides temporally of the head, axial T2 Turbo Spin Echo (TSE) scans of the brain show an increase in SNR (28-42%), reduced RF power (23.7%), and an enhanced clarity of the temporal lobes. This data suggests that not only do the scans provide clinicians with an enhanced view of the temporal lobes to further aid diagnoses, but the time required to scan is decreased along with the amount of RF power needed per scan.

Keywords

adjusted adult applicable applications around automated axial barium better body brain ceramic channel clinical coil college conditions conductivity conformable conforming constant cortex cost deuterium develop developed diameter dielectric divided drastically ears easily effective enhancement experimental fine flexibility flexible form fraction frequency funding furthermore general head health human improve improved improving indicated indicating input inside interferences journal known layers leads lobe lobes location locations logarithmic loss magnitude make material materials measured medicine methodology mixing mixture neurosurgery noise overall oxide pads part parts percentage placed powder power quality radius rapid receive recent reduced reduction removed report representative routine rule safety sections side simple slice society sought spheres spin strong subjects substitution system table tangent temporal titanate trans transmission trio turbo type underwent utilized variety volume water yang yields