Broadband Damping of Cable Modes

Ed B. Boskamp1, Masahiro Fujimoto1, Arslan Amjad1, Michael Edwards1

1GE Healthcare, Waukesha, WI, United States

Currents induced on coaxial cable shields are common in MRI due to the E fields present in the bore during the RF pulses. These cable modes can increase SAR. Cable modes can result from coupling between cable and any coil. They affect the tuning and isolation. The standard method to eliminate cable modes is a balun. We present a different method to damp the cable modes using conductive plastics. This method is broadband, and inexpensive.

Keywords

adding aligned almost amounts amping analyzer antennas apart apply around beck blue body bore broadband built cable cables capacitors carbon cause circuits clam close coax coaxial coil coils confirmed corp curve damp degree depend detuning dipole dipoles dissipated edge electric eliminating empty ends energy enough exiting feet field fields finally formed fraction gave generated give giving going half happen heat hence hole hoping impact induced inductance instead intensity interfere junction junctions kill length like local long loop made many matching material minim mode modes nodes note odes outside parallel patch pate pated perpendicular ping place placing position power proper propose pulse quiet require residing resonant rise route scale setting several shapes sheet shield shields slight solution surface tank temperature thick tissue transmit tuned understand unwanted voltage wavelength weakening ween worst