Abstract #2677

Electromagnetic Field Evaluations Inside a Body-Tissue-Simulating Cylinder Phantom Excited by an Ideal First-Order Circularly Polarized Mode

Andreas Rennings1, Le Chen1, Friedrich Wetterling2, Mark E. Ladd3, Daniel Erni1

1General and Theoretical Electrical Engineering (ATE), Faculty of Engineering, Univ. of Duisburg-Essen, Duisburg, Germany; 2Computer Assisted Clinical Medicine, Faculty of Medicine, Univ. of Heidelberg, Mannheim, Germany; 3Erwin L. Hahn Institute for Magnetic Resonance Imaging, Univ. of Duisburg-Essen, Essen, Germany

The RF field excitation homogeneity problem of MRI at high resonance frequencies is investigated on the basis of the simplest two-dimensional setup including a cylinder phantom with spatially constant but frequency-dependent material parameters and an impressed surface current distribution between the phantoms surface and a surrounding perfect electric conductor (PEC) shield. This simple arrangement has been chosen intentionally in order to reduce the set of parameters to a minimum and therefore maximize the physical insight, which allows quite general and helpful statements about the B1+ pattern quality for the widely used first-order circularly polarized (CP) excitation mode. The coefficient of variation (CoV) of the B1+ distribution is most suited as a figure of merit for the homogeneity, since it normalizes the standard deviation to the mean value of the field pattern. Only two parameters of the arrangement have been varied the excitation frequency ranging from 30MHz up to 450MHz and the diameter of the phantom ranging from a few centimeters up to 40 cm, which covers more or less every MR imaging scenario. A characteristic B1+ pattern will be introduced, which indicates the transition between a regime with the well-known central brightening and another one without.

Keywords

absorption allows another applications arrangement assisted away axis basis biological birdcage body border brightening built centimeters central characteristic characterization chosen circularly clinical coefficient computer conductivity conductor considered constant constructive contour contribution coordinate cylinder defined dependent depicted deviation diameter diameters dimensional dimensionality distribution distributions domain electrical electromagnetic engineering ensured equal equals evaluations every excitation excited extremely faculty field finite frequencies frequency full function general geometry head helpful homogeneity ideal impressed indicates inhomogeneous inside institute intentionally introduced known layered limit limiting liquid longer magnitude mark marks material maximize medicine merit mode near normalized numerically objectives pattern patterns perfect phantom physical plot polarized power practical problem proton quality quite radius ranging reasonably reduce reduced regime relatively representing required rotational scenario setup sheet shield simple simplest simulating simulation since situation smooth spatially specified square standardized statements step sufficient suited surface systems tissue took transition typical ultra univ upper variation varied wave whole widely width yielding