Abstract #2664

Fast MR Thermometry Using Phase Referenced Asymmetric Spin-Echo EPI for High Field

SUMMA25Markus Nikola Streicher1, Andreas Schfer1, Dimo Ivanov1, Laurentius Huber1, Bibek Dhital1, Enrico Reimer1, Dirk Mller1, Andre Pampel1, Robert Trampel1, Robert Turner1

1Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany

The proton resonance frequency shift MR-thermometry method is inherently very sensitive to magnetic field perturbations in time. We show an improved referenced spin-echo EPI sequence which enables paired water and fat image acquisition in 100 ms, and allows correction for almost all causes of in-vivo field drift, such as breathing. Frequency-selectivity of the spin-echo sequence is achieved by using reversed gradient amplitudes for excitation and refocusing. With an appropriate choice of parameters, the technique provides not only fat and water separation, but immediate acquisition of the reference map at the same slice position and with maximum signal amplitude.

Keywords

according accuracy achieve achieved acquisition actual affected affecting afterwards alternately always amount amplitude anterior approved around array asymmetric available better body brain cardiac channel chemical chemically close closely cognitive coil committee components condition conservative contributions coronal correct corrected correction course courses curves depending depicted deposition despite determined dirk displaced drawback drift effective either elaborate enable enough ensure ethics excitation fast field fitted fitting fluctuations formation frequencies frequency full fully ghosting good gradient gradients healthy homogeneous human immediately improved in vivo individual inferior inherently institute inversion like linear lobe local loss many maps marks minimize minor monitoring novel occipital opposite overlap park part partially peaks perturbations polarity position power practically precision probably proton proximity pulsation pulses rapid real reduced refocusing repetition representative resolution retaining reversal saturation scanner sciences select selective selectivity series several simulation skull slice slices space species spectrum spin stability step subcutaneous subjects substance substantially superior suppressed temperature thermometry unreferenced unwanted water