Abstract #2640

Active Decoupling for Prostate MR Imaging and Spectroscopy with Extended Field of View at 7T

Catalina S. Arteaga de Castro1, Ozlem Ipek1, Mariska P. Luttje1, Marco van Vulpen1, Juus Noteboom1, Peter R. Luijten1, Uulke A. van der Heide2, Dennis W.J. Klomp1

1University Medical Center Utrecht, Utrecht, Netherlands; 2Netherlands Cancer Institute, Amsterdam, Netherlands

Prostate imaging at higher magnetic fields like 7 Tesla is challenging. Even after B1 shimming the B1 strength at the prostate location remains low. Endorectal coils are used to increase the receive sensitivity at the prostate, but as transmitters they can substantially increase the B1. In contrast from receive only coils that need PIN diodes to be decoupled from the transmitters, the ERC remains tuned when combined with external coils. We use active decoupling as an alternative to PIN diode decoupling by transmitting with the ERC with an optimized amplitude and phase to counteract the coupling caused by the field of the external elements.

Keywords

achieve achieved active adjusted adjustment alternative amplitude antenna antennas applied avoid avoiding axial benefits black blocks body cancer caused challenging chosen circle close coil coils combination combined combining compartment components compromised considered contrast correspond counteract counteracted coupling decouple decoupled decoupling degrees detail diode diodes disturbance dominance double driven effective efficiency efficient element elements emulating enable enlarged even excitation expands expected extended external field fields find form generally generated in vivo inclusion inductive inserted inside institute inversion investigate like load location loop manually match matched matrix medical minimize minutes need needed needs negligible nominal optimize optimized optimizing orange particularly pelvis peter phantom position positioned power practical preferable properties propose prostate prostates radiative receive receiving reception reduced reflected remains represent requires scaled scenario sensitivity shimming situation slice spectroscopy spin spots strength strengths strong stronger substantially switching system tilted tissue torso transceiver transition transmission transmit transmitters transmitting tube tuned uniform view volunteer water white yellow zones