Quadrature Transmit Coil for Ultra High Field MRI

Zhiyong Zhai1, Michael A. Morich1

1Philips Healthcare, Cleveland, OH, United States

At ultra high fields such as 7T, B1 field distribution is dominated by the tissue dielectric effects. A transmit coil with uniform B1 remains a challenge compared to 1.5T or 3T. Here we suggest a new type of volume transmit coil for 7T and above - a birdcage structure with discontinuous (or broken) rungs. Simulations show that such a coil structure, when tuned properly, can generate similar B1 to that of a birdcage volume coil. It provides a new and more flexible way of making volume transmit coil for ultra-high fields MRI.

Keywords

advantages anatomies antenna anterior applicable array basically best better birdcage bottom break broken build central challenge channel channels chosen coil coils college comparing concepts connected continuous coronal deviation devices diameter dielectric dipole discontinuity discontinuous distance distribution divided dominated element enough even example explored feasibility feed field fields flexible generate half halves head heterogeneous human impedance improve improvement independent interleaved lateral length little loaded locations longer loops make making middle model modeled normalized nuclear orthogonal package part parts placed placing plot plotted port portions properly proposed proton quadrature radiative receive reduced regular remains resonators ring rings rung rungs separated shield shielded shimmed shimming simulations slice slices software sources space start stepped stimulus strips structure structures suggest table tissue together torso transmit transverse tuned type types ultra uniform uniformity unit various visual volume wave wavelength