Abstract #2533

Quantification Error in MREPT Due to B1 Map Inaccuracy

Jaewook Shin1, Joonsung Lee1, Jin Keun Seo2, Dong-Hyun Kim1

1Electrical & Electronic Engineering, Yonsei University, Seoul, Korea, Republic of; 2Computational Science & Engineering, Yonsei University, Seoul, Korea, Republic of

Magnetic Resonance Electrical Property Tomography (MREPT) is recently introduced to noninvasively image the distribution of electric conductivity and permittivity in the human body at the Larmor frequency. Over homogeneous regions, the relationship between the electric properties and magnetic fields can be simplified as a Helmholtz equation of B1+. In this work, the methods using B1+ maps generated by EM simulations on numerical phantoms with two tissues were proposed to investigate and evaluate how the error in the conductivity estimates using the Helmholtz equation of B1+ is related to the SNR of B1+ and the size of the tissue.

Keywords

ability acceptable added adjusted allows applied ates axial blood body boundary coil complex computation conductivity cylinders cylindrical decreased determined deviations diameter distribution dong dotted driven eases educe eight electric electrical electronic engineering equation ermined error evaluate evaluated even field fields fixed frequencies gene genera grant homogeneous human infinite inserted introduced inversely investigate location long magnitude make maps matching mate mates measured middle might model models much muscle needed noise noiseless noises noisy numerical phantoms phys posed previous program prop property proportional propose proposed prostate quad quired radius relation remove reports republic required resolution root satisfying scanner science sens sensitive shields shin simplified simulate simulated simulation slice smaller spatial square surround system tissue tissues tomography transition trials types useful varied volume wavelength wavenumber whose