Reduced Field of View Diffusion-Weighted Imaging of the Thyroid Gland Using 2D RF Pulses and Optimal B1 Reconstruction

Valentina Taviani1, Sidharta Nagala2, Andrew N. Priest3, Mary A. McLean4, Martin J. Graves3

1Department of Radiology, University of Wisconsin, Madison, WI, United States; 2Department of Otolaryngology, Addenbrookes Hospital, Cambridge, United Kingdom; 3Department of Radiology, University of Cambridge, Addenbrooke's Hospital, Cambridge, United Kingdom; 4Cancer Research UK Cambridge Research Institute, Li Ka Shing Centre, Cambridge, United Kingdom

Diffusion-weighted (DW) echo planar imaging (EPI) in the thyroid can benefit from reduced field of view (r-FOV) techniques to reduce off-resonance-induced distortion and blurring due to T2* decay. In this work, we used a 2D spatially selective RF pulse integrated into a dual spin echo DW single-shot (SS) EPI pulse sequence to perform contiguous, multi-slice, fat-suppressed, r-FOV, DW imaging of the thyroid gland at 3T. The r-FOV technique was compared with the conventional full-FOV DW-SS-EPI and the beneficial effects of an optimal B1 reconstruction strategy on the measured apparent diffusion coefficients were investigated in phantom experiments and healthy volunteers.

Keywords

accelerated account accuracy accurate achieved alleviate allows anatomy anterior apparent application applied arms array asset averaging bands basis blips body cancer carotid channel chemical coefficients coil coils containing contiguous coverage decreased designed deviation diameter differentiate diffusion dimension distortion distortions dual eddy errors evaluation except excitation fast field fields fitting full gauss gland gradient gradients graves healthy hospital identical improve in vivo included increasing induced inferior inherent inherently institute king kingdom longer magnitude malignant mapping martin matrix measured noise optimal outside overlying parallel partial peak phantom phantoms planar posterior potentially priest procedure products profile pulse pulses quantification radiology readout receive recently reconstruct reconstruction rectified reduced reduction refocused remove repetitions respectively saturation selective sense sets shot slab slew slice slow space spatial spatially spectrally spherical spin squares strength studies subject substantial suffer superior suppressed susceptibility system systematic systematically table taking thyroid tissues transmission traversal underestimated underestimation useful validate variations velocity view volume volunteers water whole