Abstract #1659

B1 Correction for Quantitative in vivo 19F Magnetic Resonance Imaging with Surface Coils

Philipp Bhm-Sturm1, Eberhard D. Pracht1, Markus Aswendt1, Nadine Henn1, Mathias Hoehn1

1In-Vivo-NMR, Max-Planck-Institute for Neurological Research, Cologne, Germany

19F MRI has recently received much attention, especially in the field of cell tracking. One of the main advantages of 19F MRI is that it is possible to quantify the number of cells. However, this only holds for the use of volume coils with a homogeneous B1 field for both transmission and reception. Here we present a workflow that allows quantitative 19F MRI with surface coils by mapping the B1 field on the 1H channel and subsequent correction of 19F data by post-processing. Our work may resolve an inherent conflict of sensitivity vs. quantification in coil design for 19F MRI.

Keywords

able absence adjusted amount animals applicability applicable appropriate assess assuming attenuation available avoid background beforehand brain cell cells certain challenging channel coil coils cologne commercially common comparing concentration concentrations concept conflict constructed control converge corrected correcting correction cost denoted density deposits designs despite dilution display division double easily echoes edges either encoding even excitation excluded experiment feasible field frequency health hemisphere highest histology homogeneous identical imperfect implantation implantations implanted in vivo included indicating inherent inhomogeneous intensity labeled linear literature long loop loss magnitude mapping maps maximize measured merged mice middle mixture mouse much neglecting neural normalize often parallel partial peak phantom post presence probably problems processing produced profile profiles progenitor proof properly protocol quantification quantitative received reception refocusing relationship rendering report resolution revealed scanner scheme sensitivity situations slice slices spectra spin spins strategies studies supported surface surgery term terms threshold together transmission tube tubes tuned turbo typical uncorrected usually vitro volume whereas yielded