Automatic Respiratory Gas Delivery Device for Noninvasive Administration of Hyperpolarized Gaseous Contrast Agents to Consciously Breathing Subjects

MAGNA25Kiarash Emami1, Hooman Hamedani1, Biao Han1, Yinan Xu1, Stephen J. Kadlecek1, Masaru Ishii2, Rahim R. Rizi1

1Radiology, University of Pennsylvania, Philadelphia, PA, United States; 2Otolaryngology - Head and Neck Surgery, Johns Hopkins University, Baltimore, MD, United States

The design of an automatic MRI-compatible respiratory gas mixing, delivery and monitoring device capable of administering a predefined tidal volume and concentration of gaseous contrast agents to consciously breathing subjects is described. This system divides the mixture of hyperpolarized gas and oxygen over several breaths for multi-breath imaging protocols or oxygen-weighted MRI, queues the subject (without the risk of over-inflating the lungs) and synchronizes image acquisition with MRI scanner. This device eliminates the need for operator supervision by responding to subjects voluntary respiratory effort, and makes it possible to maintain a respiratory pattern very similar to normal breathing.

Keywords

accuracy acquisition addition additionally adjusted adult agent agents allows animals arrangement automated automatic avoid back bags behavior breath breathing breaths buffer calibration certain close coefficient coefficients combined commencing compatible component components concentration conference continuously contrast covering defects delivery design desired developed device diagram diffusion effort eliminate enabling especially every experimentally expo extended feasibility fine flow focus freely gaseous gases healthy hods hold human humans idea identical implemented independently inhaled initial intention intervention johns king lung male manual mechanically mixed mixing mixture mixtures monitor necessitated neck need noted objectives optionally otherwise overlay oxygen path patient pattern pictures pneumatic pose predefined prescribed presence pressure problem proposed protocols prototype pulmonary readout real recent reminiscent repeatability repeated reported representative require required requirements requires resistance resp respiratory respond scanner schematic semi sense separate setup several shelf short sites sonata structure subject subjects submitted supervision system target threshold tidal tuned unmet utility various ventilation volume voluntary whereas