MRI Conditional Pacemakers: The Cooling Effect of Blood Flow in the Heart Chambers

Ramez E. N. Shehada1, Rohan More1, Sassan Rahbari1, Richard Williamson1, Ali Dianaty1

1CRMD, St. Jude Medical, Sylmar, CA, United States

MRI conditional pacemaker leads may increase in temperature primarily due to RF-induced heating; however, large segments of the leads reside in blood vessels and can benefit from the cooling effect of the surrounding blood flow. An in-vitro flow setup was developed to simulate in-vivo flow conditions around pacemaker leads and was used to determine the percent temperature reduction in the lead due to blood flow. Blood flow may reduce temperature-rises in the distal end of pacemaker leads by 83-87% and can be used to mitigate moderate heating of the distal segments of MRI conditional pacemaker leads.

Keywords

adult allow allows anatomical anatomy another apparatus around atrial bath beneficial blood body built cadaver cardiac caused chambers circulating close computed conditional conditions conductivity connecting consequent controlled cooling curves described design developed dimensions distal drop electrical ends established experimental flow harbor healthy heart heat heating implementation indicates induced initially injected injecting insulated inter investigates lead leads length long maintain maintaining male measured medical minor mitigate model models moderate monitor monitoring natural output outside pacemaker pacemakers parts patients placed positioned primarily probes protocol pump reached reaches realistic reduce reducing reduction reestablished reside resistor resistors responsible rise rises saline segments series setup several simpler simulate simulated sinking situation solution solutions specimens steady stopped studies subjects summarized summary surface table target temperature thermal third tolerated tubes typical upon various ventricular vessels vitro water wire