A Two-Compartment (Bi-Exponential) T2 Phantom with Appropriate Fat-Proton Density Correction for Validation

Burkhard Mdler1, Jrgen Gieseke2, 3

1Neurosurgery, University Bonn, Bonn, NRW, Germany; 2Philips Healthcare, Best, Netherlands; 3Radiology, University Bonn, Bonn, Germany

Validation of quantitative multi-component relaxation methods remains a challenge for the lack of suitable phantoms with known parameters. We re-facilitate the use of a fat-water phantom that uses dairy cream to validate and address the accuracy of multi-exponential analysis techniques (data acquisition and analysis). The problem of different proton densities of milk fat and water was previously not taken into consideration. Depending on the dairy cream's fat composition, primarily their content of triacylglycerides (98% of milk fat), the error in the fractional estimates can easily exceed 10-20%. We suggest an easy correction method based on partial proton densities for fat and water.

Keywords

according accuracy accurate acquisition additionally address adipose amplitudes analyzing appropriate averaging best better biggest brain cartilage causing challenges clinical common compartment compartments competing complication components comprehensively concept consideration content controlled corrected correcting correction cream decay defined definition degeneration densities density dependence depending depends depicted described desired diagnosis diagnostic diary disorders distribution double echoes enabling entry errors establish exponential field find five food fortunately fraction fractions full functions gain gained graded idea identity improvements in vivo inflammation inherently intermediate introduce introduced jersey length like long mass measured measures medium milk model molar molecule muscle neurosurgery nevertheless nominal oncology peak perfect phantom phantoms plastic popularity poses practice previously problem proportional proposed proton protons published quantification quantitative questions radiology reasons reeder regularized required requirement requires respectively separated short significantly slope spin stable stimulated studies suggested summarizes systems take taken terms tissue together tools trend tubes underneath validate validation variety various vendor volume water whipped yield yields