Abstract #0705

Imaging Without Gradients: First in vivo MR Images Using the TRASE RF Imaging Method

Jonathan C. Sharp1, Qunli Deng1, Vyacheslav Volotovskyy2, Randy Tyson1, Donghui Yin2, Richard Bernhardt2, Scott King2, Boguslaw Tomanek1

1Institute of Biodiagnostics (West), National Research Council of Canada, Calgary, Alberta, Canada; 2Institute of Biodiagnostics, National Research Council of Canada, Winnipeg, Manitoba, Canada

TRASE is an MRI acquisition method which traverses k-space using refocusing pulses applied with phase-gradient RF transmit fields. Only a single transmit channel is used, and the switched-B0 gradient system is not required. TRASE in vivo images were collected at 0.2T using a novel RF transmitter coil array, capable of producing 2-axis encoding. TRASE-encoding was used in-plane (2D), with B0-phase encoding (1D) to encode orthogonally. We collected the first in vivo TRASE images (of wrist and knee). The results are encouraging and a step towards the development of a low-cost MR imaging technology.

Keywords

acceleration achieved acquisition adapted aliasing allow alternately amplification amplifier another array arrays artifacts aspects axis becomes behavior beneficial built channel channels circuitry coil coils compatible consists console containing control controlled coronal corresponds cost costs council coverage decay described design designed dimension diode distortions duration echoes electronics eliminating encoded encoding encouraging even expected extension fairly field fields fixed focus form full future generated geometric gradient gradients identical improvements in vivo institute king knee limited limiting linear linearity long magnet magnets magnitude maintenance much national negligible next novel obtainable occurrence occurs operating optimized patient permanent phantom pixel played populations potential power preventing previously primarily principle project projection pulses reasonably reduce refocused refocusing regulations representative represented require required resolution restrictive rotated sample scheme several sharp shorter simplified since slice smear solenoid space spacing spatial spin spins square steps suited switched system though train trains trajectories trajectory transmit uniformity unlike validating visible volume west whole wider wrapping