Parallel MRI Performance Evaluation of a Novel 32-Channel Fetal Array at 1.5T

MAGNA25Ye Li1, Yong Pang1, Daniel Vigneron1, 2, Orit Glenn1, Duan Xu1, Xiaoliang Zhang1, 2

1Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, United States; 2UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco, CA, United States

In this work, we aim to evaluate parallel imaging performance of a 32 channel fetal array for fetal MRI at 1.5T. The gradient-echo image intensity of each coil element was calculated based on electromagnetic field distribution of each element. GRAPPA reconstructed images with different acceleration factors were carried out by using PULSAR toolbox. We studied the proposed flexible 32-channel array and compared with a commercial available 8-channel torso array that is routinely used in clinical fetal imaging. Our results show that the proposed fetal array improves parallel imaging performance for fetal MRI in terms of Artifact Power (AP) and image intensity.

Keywords

acceleration according acknowledgments acquisition advantages although annual array arrays artifact artifacts assuming asterisk auto available axial bioengineering biomedical block blue bottom calibration capable cardiac carried channel circularly coil coils commercial commercially complex component concepts conjugate consisted constant dash dedicated deeper defined denote denotes diminished distribution domain dramatically duration eight electromagnetic element employed encoding equation evaluate evaluation excitation excited feasibility feasible fetal field finite fitting flexible focal fold frequency gradient graduate grant grants homogeneity ignoring improve improvement improves indicate initial intensity joint linear magnetization measured meeting missing negative novel numerical numerically operation optimized pang parallel part performance phantom pixel polarized polynomial positive power previous proportional proposed pulsar pulse quantitatively radiology reconstructed reconstruction reconstructions reduced respectively routinely seed separately simulated simulation since smaller smooth space spots square supported surface susceptibility toolbox torso unable uniformly uterus utilized zoom zoomed