MOLLI T1 Values Have Systematic T2 & Inversion Efficiency Dependent Errors

MAGNA25Kelvin Chow1, Jacqueline A. Flewitt2, Joseph J. Pagano1, Jordin D. Green3, Matthias G. Friedrich2, Richard B. Thompson1

1Biomedical Engineering, University of Alberta, Edmonton, Alberta, Canada; 2Stephenson CMR Centre, University of Calgary, Calgary, Alberta, Canada; 3Siemens Healthcare, Calgary, Alberta, Canada

The MOLLI T1 mapping sequence was modeled using simulations, its accuracy investigated in phantoms and in-vivo studies, and compared to a new SAturation-recovery single-SHot Acquisition "SASHA" technique. Bloch equation simulations of MOLLI predict systematic T1 underestimation with shorter T2s and also with imperfect inversion pulse efficiency. Measured MOLLI T1s in phantoms agree well with simulations (10% underestimation of myocardium-like phantoms). Incorporating actual inversion pulse efficiency (measured in-vivo at 91%) and T2s into simulations accurately predict observed ~200ms T1 underestimation by MOLLI compared to SASHA in the myocardium. SASHA had good agreement with gold-standard T1s and showed no T2 dependence.

Keywords

absolute accumulated accuracy accurate achieved acquisition actual additional agreement assessed axial axis biomedical blood broad centric chow combine common concentrations consecutive considered consisting correction dependence dependent derived deriving described diffuse doped edema effective efficiency engineering ensure equation errors experimental explain explained explored exponential extracellular fibrosis fitting fraction full gadolinium gold good gradient green haste healthy heart heartbeats ideal implementation implementations in vivo incorporating increment incremented inherent interval intravenous inversion investigate kelvin known last like literature longer magnetization male mapping measure measured mildly minutes muscle must myocardial myocardium nearby occurs overestimated partial phantom phantoms pool post predict predicted preparation previously prior pulse quantitative readout recovery reduce repeated reported respectively routinely saturation scaling sensitivity sets short shorter shot simulate simulated simulation simulations skeletal slice smaller space spanning spin studies subjects substantial substantially suggesting systematic table timing tissue truth underestimate underestimation variants various ventricular vertical volume volunteers widely