Magnetic Resonance Encephalography Reconstruction with Magnetic Field Monitoring

SUMMA25Frederik Testud1, Jakob Asslnder1, Christoph Barmet2, Thimo Hugger1, Benjamin Zahneisen1, Klaas Prssmann2, Jrgen Hennig1, Maxim Zaitsev1

1Medical Physics, University Medical Center Freiburg, Freiburg, Germany; 2Biomedical Engineering, University & ETH Zrich, Zrich, Switzerland

In the last years field probes have been used for dynamic magnetic field monitoring in order to reduce image artifacts due to imperfect gradient coils and amplifiers, eddy currents, B0 drifts and patient breathing by taking the measured field dynamics into account in the image reconstruction. Long Magnetic Resonance Encephalography experiments suffer from B0 and gradient drifts. These can be corrected for by interleaved measurement of the k-space trajectory and the imaging experiment. This is performed in the present work with 4 </sub>1H field probes combined with 28 channels of a 95 channel head coil.

Keywords

account activity allowed amplifier array artifacts avoid beginning breathing built burg camera cardiac chain channel coil coils combination conjugate connected corrected cost derived drift drifts dwell eddy elem engineering every evolution excite experiment federal field form frame generator germ gradient grants head help in vivo leading linear long many math matrix measured measuring medical mite monitor monitoring must negligible noise nominal operated oscillations patient peak peaks phantom ported positions prob probe probes prof project pulses quired radial receive reconstruction record reduced reflects related repetition reproduced resolution respiration rich scanner sense separate series sign slab space spectrometer spectrum spherical spirals splitter square stack subtracting suffer supp taken taking temp temporal thank trajectories trajectory trigger typical variation variations varying visible volunteer years yellow