Chemical Shift Induced Phase Errors in Phase Contrast MRI

MAGNA25Matthew J. Middione1, 2, Daniel B. Ennis, 23

1Department of Radiological Sciences, Diagnostic Cardiovascular Imaging Section , University of California, Los Angeles, CA, United States; 2Biomedical Physics Interdepartmental Program, University of California, Los Angeles, CA, United States; 3Department of Radiological Sciences, Diagnostic Cardiovascular Imaging Section, University of California, Los Angeles, CA, United States

Phase contrast MRI (PC-MRI) is subject to numerous sources of error, which decrease both quantitative accuracy and clinical confidence in the reported measures. Perivascular fat surrounds most vessels and can chemically shift across the vessel wall into the lumen, thereby superposing the complex off-resonant fat signal onto the complex signal in a pixel containing flowing blood. This phase error can lead to a clinically significant over- or underestimation of net forward flow (10-20mL). A judicious choice of bandwidth and TE can reduce the chemical shift induced phase errors in PC-MRI net forward flow measurements to clinically insignificant levels (<5mL).

Keywords

absence acceleration according accuracy acquisition affected agreement aorta apparent around arteries ascending bandwidth biomedical black blood branch cardiovascular carotids changing chemical chemically choice cine clinically close collectively confidence consistently constant containing contrast contribution corrected correction dashed dataset decrease decreased decreasing define descending disease distortions eddy encoding error errors especially extended female field flatter flow flowing forward gating governed gradient gray greater heart highlight history illustrates in vivo indicate indicated induced insignificant interdepartmental judicious largely larger lead leads likely lumen made many matrix maxwell measured measures minimal minimized minimizes near numerous objectives owing paired parallel partial percent phantom physics pixel presence processing pulmonary quantitative radiological reconstructed reduce reported resolution resonant retrospective saturation section segment selection significantly slower smaller spatial spin stationary statistically step strength subject subjects subtract summarized superposing surrounds systematic systems table territories theory therein trend trio truth turbo unappreciated underestimation upward valve variance velocities velocity vessel vessels views volunteers wall years zero